THERASENSE, INC.
(Exact name of registrant as specified in its charter)
| Delaware |
94-3267373 | |
| (State of incorporation) |
(IRS Employer Identification No.) |
1360 SOUTH LOOP ROAD, ALAMEDA, CA 94502
(510) 749-5400
(Address, including zip code, and telephone number, including area code, of Registrants principal executive offices)
Securities registered pursuant to Section 12(b) of the Act:
None
Securities registered pursuant to Section 12(g) of the Act:
Common Stock, $0.001 par value
Indicate by check mark whether the registrant: (1) has filed all reports required to be filed by Section 13 or 15(d) of the Securities Exchange Act of 1934 during the preceding 12 months (or for such shorter period that the registrant was required to file such reports), and (2) has been subject to such filing requirements for the past 90 days: Yes x No ¨.
Indicate by check mark if disclosure of delinquent filers pursuant to Item 405 of Regulation S-K is not contained herein, and will not be contained, to the best of the registrants knowledge, in the definitive proxy or information statements incorporated by reference in Part III of this Form 10-K or any amendment to this Form 10-K. x
Indicate by check mark whether the registrant is an accelerated filer (as defined in Rule 12b-2 of the Securities Exchange Act of 1934): Yes x No ¨.
The aggregate market value of the 27,506,178 shares of voting stock held by non-affiliates of the registrant, computed by reference to the closing price, as reported on the Nasdaq National Market, as of the last business day of registrants most recently completed second fiscal quarter (June 30, 2002), was approximately $508,039,108. Registrant has no non-voting common equity.
There were 40,867,629 shares of the registrants Common Stock $0.001 par value, issued and outstanding as of March 1, 2003.
DOCUMENTS INCORPORATED BY REFERENCE
Certain information is incorporated by reference to the Proxy Statement for the registrants 2002 Annual Meeting of Stockholders to be filed with the Securities and Exchange Commission pursuant to Regulation 14A not later than 120 days after the end of the fiscal year covered by this Form 10-K.
| Page | ||||
| PART I |
||||
| ITEM 1. |
1 | |||
| ITEM 2. |
17 | |||
| ITEM 3. |
17 | |||
| ITEM 4. |
17 | |||
| PART II |
||||
| ITEM 5. |
MARKET FOR THE REGISTRANT'S COMMON STOCK AND RELATED STOCKHOLDER INFORMATION |
18 | ||
| ITEM 6. |
19 | |||
| ITEM 7. |
MANAGEMENT'S DISCUSSION AND ANALYSIS OF FINANCIAL CONDITION AND RESULTS OF OPERATIONS |
20 | ||
| ITEM 7A. |
44 | |||
| ITEM 8. |
44 | |||
| ITEM 9. |
CHANGES IN AND DISAGREEMENTS WITH ACCOUNTANTS ON ACCOUNTING AND FINANCIAL DISCLOSURE |
44 | ||
| PART III |
||||
| ITEM 10. |
45 | |||
| ITEM 11. |
45 | |||
| ITEM 12. |
45 | |||
| ITEM 13. |
45 | |||
| ITEM 14. |
45 | |||
| PART IV |
||||
| ITEM 15. |
EXHIBITS, FINANCIAL STATEMENT SCHEDULES AND REPORTS ON FORM 8-K |
46 | ||
| 49 | ||||
| 79 | ||||
| 81 | ||||
i
PART 1
Overview
We develop, manufacture and sell easy to use glucose self-monitoring systems that dramatically reduce the pain of testing for people with diabetes. Our first product, the Freestyle® Blood Glucose Monitoring System, received FDA clearance in January 2000, and we began selling FreeStyle in the United States in June 2000. FreeStyle utilizes patented technologies that can accurately measure glucose concentrations from a tiny 0.3 microliter sample of blood. Most of the competitive products require blood samples from 1.0 to 10.0 microliters. Our tiny sample size is easily obtained by lancing the forearm, thigh, calf, upper arm or hand and therefore avoids the pain associated with drawing a larger blood sample from the fingertip. These alternate sites are significantly less painful to lance than the traditional fingertip test site, which is more densely populated with highly sensitive nerve endings but yields the larger blood volumes required by most competitive products. Nine out of ten people in our clinical studies found using FreeStyle less painful than their current finger-stick-based system.
In June 2002, we received 510(k) clearance for and commenced sales of our FreeStyle Tracker System that incorporates the blood glucose monitoring technology from FreeStyle into a module for the Handspring Visor personal digital assistant. In July 2002, we launched our FreeStyle CoPilot System that enables people with diabetes and their health care providers to analyze and communicate glucose information using the Internet. We are also developing a Continuous Glucose Monitoring System that is intended to permit people with diabetes to accurately and discreetly measure their glucose levels on a continuous basis.
We believe that FreeStyle is well positioned to capture a meaningful share of the blood glucose self-monitoring market. The blood glucose self-monitoring market is the largest self-test market for medical diagnostic products in the world, with a size of approximately $2.0 billion in the United States and $4.0 billion worldwide. It is estimated that the worldwide blood glucose self-monitoring market will amount to $9.0 billion by 2005. We believe that FreeStyle and other products based on our proprietary technologies can expand this market by substantially reducing the pain associated with testing and thereby bring non-testers into the market and encourage under-testers to test more regularly.
Our direct sales force promotes FreeStyle in the United States to health care professionals who advise patients on the monitoring and management of their diabetes. We distribute and sell FreeStyle in the United States to the ten largest national retailers, including Walgreens, Wal-Mart and Rite Aid, through wholesalers, including Cardinal Health, McKesson, and AmerisourceBergen, and directly to end users over the telephone and through our website. In September 2000, we entered into an agreement with Disetronic Group for the exclusive distribution of FreeStyle in selected European countries. In March 2001, we obtained the CE Mark, which permits us to commercially distribute FreeStyle throughout the European Union. Disetronic commenced sales in Germany and Sweden in May 2001, and since that time has commenced sales in Norway, Finland, Austria, The Netherlands, Denmark, Switzerland, France, Italy and Belgium. In April 2001, we entered into an agreement with Nipro Corporation for the exclusive distribution of FreeStyle in Japan. An application for approval to market FreeStyle in Japan was approved by the Japanese Ministry of Health in January 2002, and Nipro launched FreeStyle in Japan in February 2002. Disetronic and Nipro are the leading suppliers of insulin pumps in Europe and Japan, respectively. Insulin pump users are typically highly motivated insulin-dependent diabetes patients who frequently test their blood glucose levels.
Market Opportunity
Diabetes
Diabetes is a chronic, life-threatening disease for which there is no known cure. Diabetes occurs when the body does not produce sufficient levels of, or fails to effectively utilize, insulin. Insulin is a hormone that
1
regulates the storage and metabolism of glucose. Glucose levels in the blood must be maintained within a specific concentration range to ensure optimal cellular function and health. Under normal conditions, the body maintains proper blood glucose levels by releasing insulin in response to increases in blood sugar.
Diabetes is typically classified as Type 1 or Type 2. Type 1 diabetes is the most serious form of the disease and is characterized by a severe lack of insulin secretion by the body. Type 1 diabetes usually occurs during childhood or adolescence, but it can occur at any age. Individuals with Type 1 diabetes require daily insulin injections to survive. Type 2 diabetes is the most common form of the disease and is characterized by the bodys inability to produce enough insulin or to properly utilize insulin. Type 2 diabetes typically occurs in adulthood. However, because of sedentary lifestyles and inappropriate diet, Type 2 diabetes is increasing in incidence among the younger population. Type 2 diabetes is initially managed with diet, exercise and oral medication. However, many people with Type 2 diabetes will eventually require daily insulin injections.
In the United States, approximately 17 million people, about 6% of the population, have diabetes, although only approximately 11 million of these people have been diagnosed with the disease. The share of the United States population diagnosed with diabetes increased 33% between 1990 and 1998, primarily due to the aging of the population, inappropriate diets and increasingly sedentary lifestyles. The most rapid onset was in adults ages 30 through 39. It is also on the rise among a younger population base, including children and teenagers. Worldwide, approximately 175 million people, about 3% of the population, have diabetes. The worldwide prevalence of diabetes is expected to increase to approximately 370 million by 2030.
Importance of Glucose Monitoring
Diabetes is the sixth leading cause of death by disease in the United States, with one death due to diabetic complications occurring every three minutes. The failure to frequently monitor and control blood glucose levels leads to severe medical complications over time, including blindness, loss of kidney function, nerve degeneration and cardiovascular disease. Diabetes is estimated to cost the United States economy over $132 billion annually, including indirect costs such as lost productivity.
The goal of glucose monitoring is to avoid the complications of diabetes by allowing patients and their health care providers to determine a treatment regimen, to monitor the effectiveness of the regimen, and to alter it as needed for better overall control of blood glucose levels. Every persons blood glucose level varies during the course of the day, depending upon factors such as diet, insulin availability, exercise, illness and stress. To successfully maintain blood glucose levels within the proper range, a person with diabetes must first measure his or her glucose level and then manage this level by adjusting insulin intake, oral medication, diet and exercise. Then the person must take additional blood glucose measurements to gauge his or her individual response to the adjustments. The more frequently people with diabetes test their blood glucose levels and track their activities and food intake, the better they will be able to understand and manage their diabetes.
Studies show that active monitoring and management of diabetes reduces the risk of associated diabetes complications. The landmark Diabetes Control and Complications Trial, or DCCT, showed that the onset and progression of eye, kidney and nerve disease in people with Type 1 diabetes can be slowed by intensive therapy to maintain blood glucose levels as close to normal as possible. The DCCT demonstrated that the risk of complications could be reduced by 76% for eye disease, 50% for kidney disease and 60% for nerve disease. Similar studies in the United Kingdom and Japan involving people with Type 2 diabetes support the conclusion of the DCCT study that actively managing blood glucose levels reduces the risk of complications associated with diabetes. People with Type 1 diabetes are encouraged to test four or more times per day, and those with Type 2 diabetes are typically expected to test two or more times per day.
Limitations of Existing Glucose Monitoring Products
Despite the proven benefits of frequent monitoring and intensive management of blood glucose levels, a significant number of people fail to test at their recommended frequency, or at all. The American Diabetes
2
Association estimates that people with diabetes test, on average, slightly more than once per day. To obtain a sample with current glucose monitoring systems, users generally are required to prick one of their fingertips with a lancing device, which typically consists of a spring-loaded needle that penetrates a measured distance into the finger. Users must then draw a sample of blood from the finger, which often requires squeezing of the fingertips. After drawing a blood sample, users generally are required to drop the blood sample on a disposable test strip or place the test strip on the blood sample. We believe that under-testing is due to the limitations of existing products including:
| | Pain. Although the fingertips are rich in capillary beds and provide a good site to obtain a blood sample, they are also more densely populated with highly sensitive nerve endings. This makes the lancing and subsequent manipulation of the finger painful. The pain and discomfort are compounded by the fact that fingers offer limited surface area, so tests are often performed on areas that are sore from prior tests. Users also suffer pain when the lance wound is disturbed during regular activities. |
| | Large Sample Size. Most competitive blood glucose meters require users to draw a sample size from 1.0 to 10.0 microliters of blood to accurately measure blood glucose levels. These larger sample sizes are difficult or impossible to obtain on sites other than the finger. Furthermore, the larger the blood sample required, the wider or deeper the lancing must be in order to reliably draw the sample. This leads to increased pain, greater likelihood of residual bleeding and longer healing time. One other blood glucose meter requires a 0.3 microliter sample size, but this sample must be obtained from the pain-sensitive fingertips. |
| | Susceptibility to Interference. The accuracy of other electrochemical-based glucose monitoring systems can be compromised in the presence of many substances commonly found in blood, such as aspirin, acetaminophen, Vitamin C and uric acid. Accuracy can also be compromised by unusually high or low levels of red blood cells. These levels can be present in infants, pregnant women, patients on dialysis, athletes and those living at high altitudes. |
| | Lifestyle Disruption. The process of measuring blood glucose levels causes significant disruption in the daily lives of people with diabetes and their families. Lancing the fingertips on infants is traumatizing to both parent and child. Obtaining large blood samples is inconvenient and may cause embarrassment in social situations, particularly for young children who are often required to be removed from class or activities to test themselves in the nurses office. |
As a result, we believe a significant market opportunity exists for a glucose self-monitoring system that combines a very small sample size requirement with the ability to test on the fingertip and other body sites. The ability to test on other body sites spares the user the pain associated with testing on the fingertip. The small sample size means the sample can be reliably obtained from other body sites and spares the user the inconvenience and social embarrassment of drawing large blood samples.
The TheraSense Solution
FreeStyle is easy to use, accurate and competitively priced. It is the only blood glucose monitoring system that combines a 0.3 microliter sample size requirement with the ability to test on other body sites and the fingertip. We believe FreeStyle also offers the following significant advantages over existing blood glucose monitoring systems:
| | Reduction in Pain. FreeStyle requires a tiny blood sample of 0.3 microliters, just a fraction of the sample size required by most other systems. The extremely small volume of blood required enables people using FreeStyle to obtain blood from their forearm, hand, thigh, upper arm or calf as well as from their fingertips as required by most other systems on the market today. Ninety percent of people in our clinical studies found using FreeStyle less painful than their current finger-stick-based systems. FreeStyle also eliminates soreness from repeated testing on a small surface area. |
| | Better Performance. FreeStyles proprietary measurement technology is extremely accurate, operates over a broad temperature range and is unaffected by common interfering substances, such as aspirin, |
3
| acetaminophen, Vitamin C and uric acid. It is also unaffected by unusually high or low levels of red blood cells. The tiny blood sample required by FreeStyle can be reliably obtained from sites other than the fingertip. |
| | Improved Quality of Life. The combination of a smaller sample size and off-fingertip testing enabled by FreeStyle significantly reduces residual bleeding. This reduces the embarrassment of testing felt by some people with diabetes and affords them more discretion in testing. The pain and awkwardness of publicly obtaining large blood samples has deterred some people with diabetes from testing frequently enough to properly manage their disease. |
Our Strategy
Our objective is to be a leading provider of innovative glucose self-monitoring products that reduce the pain of testing, are easy to use, accurate, cost effective and improve the lives of people with diabetes. To achieve this objective, we are pursuing the following business strategies:
| | Establish FreeStyle as a leading blood glucose self-monitoring device. We are creating awareness of the advantages of FreeStyle in the United States among health care professionals and people with diabetes. We do this through advertising, extensive retail distribution and our sales force. We believe an increased awareness of FreeStyles less painful, more discreet and reliable process will lead many current testers to switch to FreeStyle. In addition, we believe we can expand the market to those people who have been diagnosed with diabetes but are currently not testing, as well as increase testing frequency for those who are under-testing. |
| | Maintain and enhance retail distribution. We currently have authorized shelf space with the ten largest chain drug stores, the three largest mass market retailers and the three largest supermarket retailers in the United States. These retailers represent over 20,000 pharmacy outlets in the United States. We plan to continue to expand FreeStyles availability within these distribution channels through our national accounts sales representatives that interact with the retailers at the corporate and district level and geographically dispersed sales representatives that call on health care professionals including pharmacists. |
| | Focus on our core competencies. We plan to continue to focus our internal resources on our core competencieselectrochemistry and sensor manufacturing technologies. Consequently, we have entered into strategic relationships to enhance speed to market and cost effectiveness for those business functions not included in our core competencies. For example, we have a strategic relationship with Flextronics International, which is currently manufacturing our meters and assembling our FreeStyle System kits. Through these relationships, we believe that we will be able to quickly and efficiently build infrastructure and services needed to meet anticipated market demand. |
| | Provide high quality customer service. We provide all of our customers with easy, comprehensive access to our products and services through the use of sophisticated software systems and an educated and caring customer service team. Our approach is to partner with a service organization while maintaining a small team of in-house service specialists to monitor quality. We offer customer service 24 hours per day, seven days per week with access to dedicated representatives via telephone or the Internet. In addition, we use the Internet to enable customers to purchase our products online, enhance awareness of our products, establish e-mail management, facilitate loyalty programs and provide product support and training. |
| | Expand International Distribution. We intend to expand our international sales of FreeStyle and enter new global markets primarily through relationships with established health care companies that have developed distribution channels. The Disetronic Group is our exclusive distributor of FreeStyle in selected European countries. Disetronic has commenced sales in Germany, Sweden, Norway, Finland, Austria, The Netherlands, Denmark, Switzerland, France, Italy and Belgium. In April 2001, we entered into an agreement with Nipro Corporation for the exclusive distribution of FreeStyle in Japan. Nipro launched FreeStyle in Japan in February 2002. Disetronic and Nipro are the leading suppliers of insulin |
4
| pumps in Europe and Japan, respectively. Insulin pump users are typically highly motivated insulin-dependent diabetes patients who frequently test their blood glucose levels. |
| | Leverage our proprietary technology platform. We intend to leverage our proprietary electrochemical sensor technologies to develop new glucose monitoring products. We are currently developing a Continuous Glucose Monitoring System intended to continuously measure and display a persons glucose levels in real time for up to three days. We are also expanding our current FreeStyle product family by developing enhanced versions of FreeStyle. FreeStyle Tracker is a module for the Handspring Visor personal digital assistant that enables it to act as a glucose monitor and sophisticated diabetes management system. We obtained FDA clearance for FreeStyle Tracker and commenced sales in June 2002. |
Our Products
FreeStyle Blood Glucose Monitoring System. Our initial product, the FreeStyle blood glucose monitoring system, received FDA clearance in January 2000 for use on the forearm and fingers. We began selling FreeStyle in the United States in June 2000. In December 2000, we received FDA clearance that permits FreeStyle to be used on the thigh, calf, upper arm and hand. This represents the broadest array of off-finger testing sites cleared by the FDA. The FreeStyle System kit includes a FreeStyle meter, an initial supply of 10 proprietary disposable FreeStyle test strips, a FreeStyle lancing device, an initial supply of 10 disposable FreeStyle lancets, FreeStyle control solution and instructional materials. We also sell additional supplies of disposable FreeStyle test strips in quantities of 25, 50 and 100 and additional supplies of disposable FreeStyle lancets in quantities of 100.
| | FreeStyle meter. The FreeStyle meter contains a large display screen to read test results, a slot where the test strip is inserted to get a blood glucose reading, and buttons to change the calibration code and review results in the system memory. It also contains a data port for transmitting information to our FreeStyle CoPilot web-based data management system and FreeStyle Connect data management software. The ergonomically designed meter fits easily in the hand and weighs 2.1 ounces. The meter displays blood glucose results in a range of 20 to 500 mg/dl. Once the sample is acquired, the meter takes about 15 seconds to display the result. The meter has the ability to store the last 250 blood glucose test results and to display a 14-day average blood glucose level. |
| | FreeStyle test strips. FreeStyle test strips are proprietary disposable sensors that are used with the FreeStyle meter to measure blood glucose levels. The test strips are clearly marked to indicate proper placement in the meter. Inserting the test strip into the meter activates the system and either side of the test strip can be used for measurement. The FreeStyle meter beeps one time when sufficient blood has been drawn into the test strip and beeps two times when the test is complete. Our proprietary FreeStyle test strips may only be used with our FreeStyle meter and FreeStyle Tracker System. |
| | FreeStyle lancing device and lancets. The FreeStyle lancing device is designed specifically to make blood sample acquisition reliable and convenient. It requires no mechanical or vacuum assistance to draw blood. The lancing device offers five adjustable depth settings to allow for comfort and adequate sample size. Although FreeStyle lancets are available, other standard lancets are compatible with our system. It is recommended that a new, sterile lancet be inserted into the lancing device every time a test is administered. The reduction in pain from FreeStyle is attributable to the lancing site and the small sample size required, not the type of lancing device or lancet. |
| | FreeStyle control solution. The FreeStyle control solution contains a fixed amount of glucose that may be used periodically to ensure the FreeStyle System is functioning correctly and users are following correct testing procedures. |
FreeStyle Tracker Diabetes Management System. We received FDA clearance for the FreeStyle Tracker System and commenced sales in June 2002. The FreeStyle Tracker System is a diabetes management system that combines the blood glucose monitoring technology of our FreeStyle Blood Glucose Monitoring System with the computing power of the HandSpring Visor personal digital assistant, or PDA. The FreeStyle Tracker System
5
enables patients to test for glucose levels and get a read-out on the PDA screen, graph and chart the results over time, review food lists to track their carbohydrate intake and create reminders about testing or dietary choices. For doctors, the FreeStyle Tracker System provides a time-stamped progression of the patients glucose levels as well as patient-entered events and data that affect their diabetic health to make better diagnoses and recommendations for the patients diabetes care. Patients can use the Hotsync® synchronization function of the PDA to download information from the FreeStyle Tracker System to our FreeStyle CoPilot System and FreeStyle Connect Software.
FreeStyle CoPilot Web-Based Data Management System. We launched the FreeStyle CoPilot System in July 2002. The FreeStyle CoPilot System is a free web-based service that we offer to users of blood glucose monitoring products and their permitted health care providers. Users can upload their blood glucose test results and other health-related information from their FreeStyle meter (through a FreeStyle Connect data cable) or FreeStyle Tracker System to the FreeStyle CoPilot or users can manually enter this information into the FreeStyle CoPilot System. The information can then be transferred into personalized reports focusing on integrated issues of diabetes care, including glycemic control, hypertension, hyperlipidemia, and diet. Users can grant permission to their health care providers to access the users information on the FreeStyle CoPilot System.
FreeStyle Connect Data Management Software. We received FDA clearance for the FreeStyle Connect software in May 2000 and we launched the FreeStyle Connect software in December 2000. FreeStyle Connect is a data management software product that downloads data from FreeStyle to a personal computer and displays glucose values in eight different statistical reports, including the number of blood glucose values above, within, and below a given target range. The FreeStyle meter stores up to 250 glucose values each with time and date. This data allows FreeStyle customers and their health care providers to appropriately adjust customers diet, exercise and medication to improve and maintain their health.
Products Under Development
Continuous Glucose Monitoring System. We are developing a continuous monitoring device that will utilize a disposable, miniaturized electrochemical sensor that can be inserted under the skin by the user utilizing a spring-loaded insertion device. This sensor system will enable users to continuously measure and display glucose levels and store the results for further analysis by the user or health care providers. This product is intended to act as a substitute for current glucose self-monitoring devices. The increased number of glucose readings will allow people with diabetes to more effectively adjust insulin, oral medication, diet and exercise, which should result in significantly improved health outcomes for people with diabetes. The Continuous Glucose Monitoring System is being designed to offer people with diabetes the following benefits:
| | accurate and discreet measurement of glucose levels on a continuous basis, and display of trends; |
| | elimination of the anxiety of not knowing glucose levels between periodic measurements; |
| | minimally invasive insertion procedure; |
| | comfort during use; |
| | warnings against dangerously high or low glucose levels, even while sleeping; and |
| | ability to improve health through intensively managed therapy from continuous glucose information. |
We believe each sensor used with our system will provide up to three days of continuous glucose measurement. The accuracy and precision of our Continuous Glucose Monitoring System will be dependent on the initial calibration. Therefore, our system will have a built-in FreeStyle meter that will allow for accurate and convenient calibration using FreeStyle test strips. The integrated calibration will eliminate the risk of human error during data entry. The display unit, which can be worn like a pager, will translate the sensors information into a numerical value and periodically, or on demand, display the glucose level and trend. This information will allow users to determine their blood glucose value and whether it is rising, falling or remaining stable. The sensor
6
system is designed to communicate to the wireless display unit within a 10-foot range, so it can be conveniently worn on a belt, carried in a purse or left on a bed stand at night.
We have commenced a home-use study to evaluate the safety and efficacy of our Continuous Glucose Monitoring System. Following completion of this study, we will submit a premarket approval application. The premarket approval process requires considerably more data and FDA review time than the 510(k) clearance process that was applicable to FreeStyle. The premarket approval process generally takes between one and three years from completion of an application, but may take longer. However, achieving a completed application is a process that may take numerous clinical trials and require filing of amendments over time. Therefore, even if the Continuous Glucose Monitoring System is successfully developed, it may not be commercially available for a number of years.
Our Sensor Technologies
We have developed two proprietary miniaturized electrochemical sensor technologies. The first, NanoSample technology, is used in our FreeStyle System and FreeStyle Tracker System. The second, Wired Enzyme chemistry, is used in our Continuous Glucose Monitoring System under development.
NanoSample Technology. NanoSample technology enables FreeStyle to measure glucose levels in blood samples of only 0.3 microliters, a fraction of the sample size required by most competitive products. We have pioneered techniques to obtain accurate, reliable and fast responses when measuring glucose in sub-microliter sample sizes. This technology allows us to measure the total electrical charge generated by the reaction of all of the glucose in the sample, a process referred to as coulometry. In contrast, the most advanced competitive products generally determine glucose levels by taking a measurement of the current generated by the sensor at a point in time, a process referred to as amperometry. Amperometry usually requires the use of a larger blood sample to achieve accurate results. Use of coulometry substantially eliminates some of the errors frequently associated with amperometry, such as dependence of sensor output on temperature and potential interference from commonly found substances in the blood, such as aspirin, acetaminophen, Vitamin C and uric acid, which can distort the glucose measurement.
Wired Enzyme Chemistry. Our Wired Enzyme chemistry is allowing us to develop miniaturized, self-insertable, biocompatible, disposable sensors. We are currently using this technology to develop our Continuous Glucose Monitoring System. Our Continuous Glucose Monitoring System sensor, which will be inserted under the skin by the user, will react with the glucose near or at the insertion site to produce an electrical signal that enables glucose concentration measurement. We believe our technology will successfully address the core technical issues that have limited the performance of other implantable glucose sensors, including oxygen dependence and interference from commonly found substances in blood, such as aspirin, acetaminophen, Vitamin C and uric acid. We also believe our system will be calibrated easily and accurately.
Marketing and Sales
United States. Our marketing and sales program is intended to generate awareness of FreeStyle and penetrate and expand the glucose self-monitoring market. The sales force includes sale representatives who promote FreeStyle to the health care professionals who strongly influence the health care decisions made by people with diabetes, a group which includes endocrinologists, certified diabetes educators, pharmacists and internal medicine physicians. The primary goal of our sales representatives is to educate and train health care professionals on the benefits of our products. We also provide these health care professionals with free samples of our products. There are also members of our sales force dedicated to serving retail and managed care accounts at the corporate and district level. In addition, our sales force promotes FreeStyle and monitors stocking levels with retail outlets at the individual store level. We believe that our strategy of selling through our own direct sales force is an important factor in achieving market penetration.
7
Our direct-to-consumer advertising campaign is aimed at health care professionals, people with diabetes and people who know people with diabetes. Our belief is that pain, reliability and quality of life issues are so important in glucose testing that they are recognized and understood not only by people with diabetes, but also by their co-workers, friends, and families, each of whom will be willing to tell others. To further generate awareness and penetrate the market, our sales and marketing organization provides a wide range of programs, support materials and events that support our national sales force. These include public relations efforts, product training, conference and trade show attendance, and educational and promotional literature.
We primarily sell our products through retail pharmacies. We sell our products directly to national retail pharmacies and supply other retail pharmacies through wholesalers. We also sell to durable medical equipment suppliers and directly to end users through phone orders and our website. Although there is substantial competition from existing products, the consolidation of the retail industry has allowed us to concentrate our sales efforts. The following is a list of our top five retailers and top five wholesalers, ranked by dollar volume of sales for the year ended December 31, 2002:
| Retailers |
Wholesalers | |
| Walgreens |
Cardinal Health | |
| Wal-Mart |
McKesson | |
| Rite Aid |
AmeriSource Bergen | |
| CVS |
Peytons Northern | |
| Eckerd |
QK Healthcare |
International. We intend to expand our international sales efforts for our FreeStyle blood glucose monitoring system and enter new global markets by establishing relationships with international partners who have established relationships with healthcare professionals and developed distribution channels. In March 2001, we obtained the CE Mark, which permits us to commercially distribute FreeStyle throughout the European Union. In September 2000, we entered into an agreement with Disetronic Group for the exclusive distribution of FreeStyle in Germany, Switzerland, Denmark, Austria, Sweden, Finland, Norway and the Netherlands. In February 2002, this agreement was amended to add France, Italy and Belgium to Disetronics exclusive distribution territory. In April 2001, we entered into an agreement with Nipro Corporation for the exclusive distribution of FreeStyle in Japan. Disetronic and Nipro are the leading suppliers of insulin pumps in Europe and Japan, respectively. Insulin pump users are typically highly motivated insulin-dependent diabetes patients who frequently test their blood glucose levels.
Under the terms of the Disetronic Group agreement, Disetronic has exclusive responsibility for sales, marketing and customer service in its territory in Europe. We may terminate the agreement if Disetronic does not meet specified minimum purchase requirements or for any other reason upon prior written notice. Disetronic is also entitled to market FreeStyle to Disetronics pump users in North America. The term of the Disetronic agreement ends in December 2006. Disetronic commenced sales in Germany and Sweden in May 2001. Since that time they have commenced sales in Norway, Finland, Austria, The Netherlands, Denmark, Switzerland, France, Italy and Belgium. In January 2003, we amended our agreement with the Disetronic Group. Pursuant to the amendment, Disetronic Injection Systems will take over distribution of our FreeStyle blood glucose monitoring system from Disetronic Medical Systems. We received $15.0 million pursuant to the amendment, and we will recognize the $15.0 million over the remaining term of the agreement, which expires in December 2006. Although Disetronic Injection Systems will be subject to reduced minimum purchase requirements, Disetronic Injection Systems has committed to increase the sales force currently selling the FreeStyle system in Europe by 50%.
Under the terms of the Nipro Corporation agreement, Nipro will have exclusive responsibility for sales, marketing and customer service in Japan. We may terminate the agreement if Nipro does not meet specified minimum purchase requirements. The initial term of the Nipro agreement ends in April 2006. FreeStyle received
8
regulatory approval for marketing in Japan in January 2002, and Nipro launched FreeStyle in Japan in February 2002.
In addition to the countries covered by the distribution agreements with Disetronic Group and Nipro Corporation, as of March 1, 2003, FreeStyle products are sold in 20 other countries through local distributors.
Distribution. To establish a worldwide distribution capability for end users, health care professionals and retail customers, we have established relationships with expert distribution partners. For retail order management and shipping of our FreeStyle System kit and other products, we have entered into an exclusive services agreement with UPS Supply Chain Management, a division of UPS Global Logistics that specializes in providing outsourced distribution services for large pharmaceutical and medical device companies. The initial term of this agreement ends in March 2005. We may terminate this agreement prior to March 2005, subject to payment of a termination fee. UPS Supply Chain Management has an extensive network of distribution centers and a sophisticated order management and product tracking system. UPS Supply Chain Management also manages our billing process. Our relationship with UPS Supply Chain Management allows us to meet shipment, delivery and billing expectations while minimizing our internal infrastructure requirements.
Customer Service. We provide customer service 24 hours per day, seven days per week through ICT Group. This service is transparent to the caller and provides a standard of service expected in the industry. This relationship with ICT Group provides customer service, technical support, a help desk and order processing. ICT Group is an international telemarketing and e-support company, with a medical marketing division which has developed a special facility and dedicated customer care agents for us. ICT Groups agents have the systems capability to handle large volumes of our customer contacts at any time, both over the phone or through our web site. We select and train the ICT Group agents who work on our account, as well as maintain in-house customer service personnel that monitor quality. Our non-exclusive contract with ICT has an initial term of three years, ending in April 2003, although it can be terminated by either party without cause upon 120 days notice. We are currently negotiating an amendment with ICT to, among other things, extend the term of the contract.
Manufacturing
The primary components of the FreeStyle System kit are the FreeStyle meter, FreeStyle disposable test strips, the FreeStyle lancing device, FreeStyle disposable lancets and FreeStyle control solution. The primary components of the FreeStyle Tracker System kit are the FreeStyle Tracker module that is inserted into the HandSpring Visor PDA, FreeStyle disposable test strips, the FreeStyle lancing device, FreeStyle disposable lancets and Freestyle control solution. We manufacture the FreeStyle test strips and contract with third parties for the manufacture of the other FreeStyle products. These contract manufacturing relationships minimize our capital investment, help control costs and allow us to compete with larger volume manufacturers of blood glucose self-monitoring systems.
We manufacture the FreeStyle test strips at our facility in Alameda, California. We have developed a manufacturing process for the test strips that we believe is robust, cost effective and scaleable to meet higher volumes. The test strip is composed of chemicals, adhesive and a printed polyester similar to the material used in credit cards.
Flextronics International assisted us in the design of our FreeStyle meter and FreeStyle Tracker module. Flextronics is responsible for manufacturing the FreeStyle meter in China and assembling the FreeStyle System kits in San Jose, California. Flextronics is also responsible for manufacturing the FreeStyle Tracker module and assembling the FreeStyle Tracker System kits in San Jose, California. Flextronics has over 13 years of experience building blood glucose meters, and has facilities in Asia, Europe and the Americas. Flextronics has demonstrated strong process control and knowledge of just-in-time and total quality management techniques and has software tools to handle product tracking. We have an on-site manager at Flextronics in San Jose who is responsible for the day-to-day interface with Flextronics. Production release to finished goods inventory is done through our
9
quality assurance department. Our contract with Flextronics expires in November 2004, and is renewable annually thereafter. Either party may terminate this contract for any reason upon one years prior written notice to the other.
Facet Technologies LLC, a wholly-owned subsidiary of Matria Healthcare, assisted us in the design of the FreeStyle lancing device and we have agreed to purchase the FreeStyle lancing devices and lancets exclusively from Facet until June 1, 2007. Facet is a leading supplier of lancing devices and lancets, including our lancets. Our FreeStyle lancing device can also use conventional lancets, which are widely available.
Each of the production processes utilized in the manufacture of our products has been verified and validated, as required by the FDAs quality system regulations. As a medical device manufacturer, our manufacturing facility and the facilities of our suppliers, such as Flextronics and Facet Technologies, are subject to periodic unannounced inspection by regulatory authorities and these operations may undergo compliance inspections conducted by the FDA and corresponding state agencies.
Intellectual Property
We rely on a combination of copyright, patent, trademark, trade secret and other intellectual property laws, nondisclosure agreements and other measures to protect our proprietary rights. As of March 1, 2003, we had 37 issued U.S. patents and had numerous additional U.S. patent applications pending. We believe it will take up to five years, and possibly longer, for some of these U.S. patent applications to result in issued patents. We have also filed foreign patent applications on our technology. Our issued patents expire between November 2010 and October 2019. The issued patents cover, among other things:
| | the designs of our FreeStyle meter and FreeStyle strip and FreeStyle lancing device products; |
| | lancing devices of the type sold with our FreeStyle blood glucose monitoring system; |
| | aspects of glucose measurement in small sample volumes using electrochemical sensors, such as those using coulometry, those having certain fill detection features, and those having certain sensor chemistries; |
| | our Wired Enzyme chemistry; |
| | a one-point calibration method useful in our Continuous Glucose Monitoring System; |
| | manufacturing processes for sensors useful in our Continuous Glucose Monitoring System; |
| | certain sensing and electronic components and methods of use for our Continuous Glucose Monitoring System; |
| | an electrochemical affinity assay system; |
| | a biological fuel cell; and |
| | electrochemical methods for verifying amplification of nucleic acids. |
We have obtained registrations for the trademark TheraSense in the U.S., Canada, Europe and Japan and have applied to register TheraSense in numerous other jurisdictions as well. We have obtained registrations for the trademark FreeStyle in the U.S. and Japan and have applied to register FreeStyle in numerous other jurisdictions including Canada and Europe.
In addition to developing our own technology, we have entered into several license agreements. We have acquired rights to patents from the University of Texas at Austin developed by Professor Adam Heller, a co-founder of our company, and his collaborators. We also fund ongoing research at the University of Texas at Austin in the field of biosensors and enzyme electrodes, and we are the licensee of resulting inventions. We have also obtained non-exclusive, worldwide licenses to specific patents owned by Asulab SA and Inverness Medical
10
Innovations, Inc. Each of these licenses grants us the right to use the licensed patents to make and sell diagnostic devices for diabetes monitoring that contain the licensed technology. We pay for these licenses through a combination of fixed payments and royalties on sales of covered products. Each of these licenses continues until expiration of the licensed patents.
Research and Development
Our research and development efforts are currently focused on developing further enhancements to our FreeStyle products as well as developing our Continuous Glucose Monitoring System. Our research and development staff has extensive experience in the glucose monitoring industry and has been instrumental in technology development and commercialization of glucose monitoring products. Research and development expenses, including clinical and regulatory expenses, were $12.0 million in 2000, $16.1 million in 2001 and $20.3 million in 2002. We expect research and development expenses to continue to increase as we seek to enhance our existing products and develop additional products.
We also fund biosensor and enzyme electrode research under a Sponsored Research Agreement with the University of Texas at Austin. We have specific rights with regard to inventions resulting from the research. The research is currently under the direction of Professor Adam Heller and is focused on improvements to implantable glucose sensors and on extension of the Wired Enzyme technology for the measurement of other biochemicals. This agreement continues on a year-to-year basis unless otherwise agreed by the parties and so long as the University has received sponsored research funds from us in the prior six-month period. We fund such research on a cost plus reasonable overhead basis.
Competition
The medical device industry is subject to intense competition. Four companies, Roche Diagnostics, LifeScan, Inc., a division of Johnson & Johnson, Bayer Corporation and MediSense, a division of Abbott Laboratories, currently account for approximately 90% of the worldwide sales of blood glucose self-monitoring systems. All of these competitors products use a meter and disposable test strips to test blood obtained by lancing the finger and, in some cases, other body sites. All of these competitive products require significantly larger blood samples than required by FreeStyle. Two products cleared for use on the finger and forearm offer a faster test time than FreeStyle, once the required sample has been obtained, and operate over a broader temperature range. Also, a new competitor has recently launched a blood glucose monitoring system that claims the same sample size requirement as the FreeStyle System. However, this product is only cleared for use on the fingertip.
In addition, other companies are developing and/or marketing minimally invasive or noninvasive glucose monitoring devices and technologies that could compete with FreeStyle and our proposed Continuous Glucose Monitoring System. There are also a number of academic and other institutions involved in various phases of our industrys technology development. Many of our competitors have significantly greater financial and human resources than we do. At this time, there is one cleared and one approved product for continuous glucose monitoring, neither of which is presently approved as a substitute for current glucose self-monitoring devices. The continuous glucose monitoring system developed by Medtronic MiniMed that has been cleared by the FDA, includes an implantable sensor that measures and stores glucose values every five minutes, for a period of two to three days. The Medtronic MiniMed system is not a consumer product, rather, it is a physician product. The sensor is required to be implanted by a physician, and the results of the data aggregated by the system can only be accessed by the physician, who must extract the sensor and download the results for viewing using customized software. The second product for continuous glucose monitoring that has been approved by the FDA, developed by Cygnus Inc., is worn on the wrist like a watch and can take glucose readings as frequently as every ten minutes for up to twelve hours at a time.
11
We believe that the principal competitive factors in our market include:
| | improved outcomes for people with diabetes through less painful, more accurate and more convenient testing methods; |
| | technological leadership in features and reliability; |
| | equivalent reimbursement from third-party payors among competitive brands; |
| | customer focus and service; |
| | effective marketing and distribution; |
| | acceptance by health care professionals; and |
| | speed to market. |
Government Regulation
Our products are medical devices subject to extensive regulation by the FDA and other regulatory bodies. FDA regulations govern, among other things, the following activities that we or our partners perform and will continue to perform:
| | product design and development; |
| | product testing; |
| | product manufacturing; |
| | product labeling; |
| | premarket clearance or approval; |
| | advertising and promotion; and |
| | product sales and distribution. |
FDAs Premarket Clearance and Approval Requirements
Unless an exemption applies, each medical device we wish to commercially distribute in the United States will require either prior 510(k) clearance or prior premarket approval from the FDA. The FDA classifies medical devices into one of three classes. Devices deemed to pose lower risk are placed in either class I or II, which requires the manufacturer to submit to the FDA a premarket notification requesting permission for commercial distribution. This process is known as 510(k) clearance. Some low risk devices are exempted from this requirement. Devices deemed by the FDA to pose the greatest risk, such as life-sustaining, life-supporting or implantable devices, or devices deemed not substantially equivalent to a previously cleared 510(k) device are placed in class III, requiring premarket approval.
510(k) Clearance Pathway. To obtain 510(k) clearance, we must submit a premarket notification demonstrating that the proposed device is substantially equivalent to a previously cleared 510(k) device or a device that was in commercial distribution before May 28, 1976 for which the FDA has not yet called for the submission of premarket approval applications. The FDAs 510(k) clearance pathway usually takes from four to twelve months from the date the application is completed, but it can take significantly longer.
Blood glucose testing systems have generally qualified for clearance under 510(k) procedures. We received 510(k) clearance for FreeStyle in January 2000 for use on the fingers and forearm. In May 2000, we also obtained 510(k) clearance for FreeStyle Connect, our data management system that enables downloading of blood glucose data stored in a users FreeStyle monitor to a personal computer for use by the user or his or her health care provider. In December 2000, we received 510(k) clearance allowing us to promote FreeStyle for use
12
on the thigh, calf, upper arm, and hand, in addition to the fingers and forearm. In December 2001, we received 510(k) clearance for certain labeling changes that we made to FreeStyle. In June 2002, we received 510(k) clearance for our FreeStyle Tracker system, our system that incorporates the blood glucose monitoring technology of FreeStyle into a module for the HandSpring Visor handheld computer.
After a device receives 510(k) clearance, any modification that could significantly affect its safety or effectiveness, or that would constitute a major change in its intended use, will require a new 510(k) clearance or could require premarket approval. The FDA requires each manufacturer to make this determination initially, but the FDA can review any such decision and can disagree with a manufacturers determination. If the FDA disagrees with a manufacturers determination, the FDA can require the manufacturer to cease marketing and/or recall the modified device until 510(k) clearance or premarket approval is obtained. We have modified aspects of FreeStyle since receiving regulatory appr